FSSAI Food Analyst and Junior Analyst Examinations · India · Food Safety and Standards Authority of India
Original practice questions written to the published syllabus · not past papersQuality systems, statistical process control, sampling inspection and running an accredited dairy laboratory.
In the exams: ICAR AICE subject 34 Unit 9, subject 35b Unit 1 and subject 36 Unit 6
How our questions for FSSAI Food Analyst and Junior Analyst Examinations spread across the units of Quality Assurance, Statistical Quality Control & Laboratory Management. Difficulty: 6 easy, 17 medium, 7 hard.
| ISO/IEC 17025 and laboratory accreditation | 11 | |
| Laboratory set-up, safety and chemical management | 10 | |
| Laboratory audits and proficiency testing | 9 |
Choose an answer, then open the working. Each explanation says why the right answer is right and why the others are not.
How many standard sections does a GHS safety data sheet contain?
A. 16
The GHS safety data sheet has sixteen sections, from identification and hazard identification through handling, exposure controls and disposal to regulatory and other information.
Further reading: UN Globally Harmonized System (GHS)
When preparing a dilute acid in the laboratory, the safe practice is to:
A. Add the acid slowly to the water with stirring
Adding acid to water lets the large volume of water absorb the heat of dilution. Adding water to concentrated sulphuric acid can boil locally and spatter acid.
Further reading: Standard laboratory safety practice
A plant laboratory is asked to culture and confirm Brucella from raw milk samples. What is the most appropriate response?
C. Send the work to a laboratory with the higher containment it requires
Brucella is a well-known cause of laboratory-acquired infection, and work with its cultures needs higher biosafety containment than routine indicator counts. Fume cupboards protect against chemical vapours, not aerosols of pathogens; preservatives do not make culture safe.
Further reading: WHO Laboratory Biosafety Manual
In ISO/IEC 17025:2017, impartiality and confidentiality are addressed in which clause?
C. Clause 4, general requirements
Clause 4 (general requirements) covers impartiality and confidentiality. Clause 6 covers personnel, facilities and equipment; clause 7 the testing process; clause 8 the management system.
Further reading: ISO/IEC 17025:2017
Under ISO/IEC 17025:2017, what is 'option B' for the management system?
D. Operating an ISO 9001 system that also supports the laboratory requirements
Clause 8 lets a laboratory meet the management-system requirements either directly (option A) or through a system run to ISO 9001 that supports the technical requirements (option B). Internal audit is required either way, and ISO does not audit laboratories.
Further reading: ISO/IEC 17025:2017
In measurement uncertainty, what is a Type B evaluation?
C. Evaluation by means other than statistics of repeated measurements
Type A components are evaluated statistically from repeated observations; Type B components come from other information such as calibration certificates, tolerances and published data. The labels say nothing about the type of test.
Further reading: JCGM 100:2008 (GUM)
Using the convention LOD = 3 × s(blank) and LOQ = 10 × s(blank), what are the LOD and LOQ when the standard deviation of blank results is 0.4 µg/kg?
B. LOD 1.2, LOQ 4.0 µg/kg
LOD = 3 × 0.4 = 1.2 µg/kg; LOQ = 10 × 0.4 = 4.0 µg/kg. Other conventions exist, which is why the stem states this one. Swapping them or using 1× and 3× gives the wrong pairs.
Further reading: Eurachem, The Fitness for Purpose of Analytical Methods
Organisations that run proficiency-testing schemes are accredited against which standard?
A. ISO/IEC 17043
PT providers are accredited to ISO/IEC 17043. ISO/IEC 17025 is for testing and calibration laboratories, ISO 9001 for quality management systems and ISO 19011 gives guidance on auditing.
Further reading: ISO 13528 and ISO/IEC 17043
What is the difference between a correction and a corrective action?
B. A correction fixes the instance; corrective action removes the cause
Correction deals with the immediate nonconformity (for example repeating a test); corrective action finds and eliminates the root cause so it does not recur, and its effectiveness is checked.
Further reading: ISO 9000:2015 and ISO 9001:2015
Routine mid-infrared milk analysers used for payment are kept accurate mainly by:
D. Regular calibration against reference-method values on calibration milks
MIR instruments are indirect methods calibrated against reference methods (such as Röse-Gottlieb for fat and Kjeldahl for protein) using calibration milks, with slope and intercept checks (see ISO 9622 | IDF 141 and ISO 8196 | IDF 128). Lamp replacement or day-to-day comparison alone does not ensure accuracy.
Further reading: ISO 9622 | IDF 141; ISO 8196 | IDF 128