FSSAI Food Analyst Examination (and Junior Analyst Examination) · India · Food Safety and Standards Authority of India
Original practice questions written to the published syllabus · not past papersControlling quality in the plant and in the laboratory: quality systems and TQM, control charts and process capability, acceptance sampling and two- and three-class sampling plans, laboratory quality management and ISO/IEC 17025 accreditation, method validation and proficiency testing.
In the exams: ICAR AICE 35.a Unit 4 (quality systems, control charts, acceptance sampling) and 35.b Units 1 and 3 (ISO 17025, NABL accreditation, 2- and 3-class sampling plans); food analyst examinations
How our questions for FSSAI Food Analyst Examination spread across the units of Food Quality Assurance, Statistical Process Control & Laboratory Accreditation. Difficulty: 10 easy, 27 medium, 2 hard.
| Method validation: accuracy, precision, LOD, LOQ and recovery | 8 | |
| Laboratory quality management and ISO/IEC 17025 | 7 | |
| Measurement uncertainty | 7 | |
| Laboratory accreditation bodies and their role | 7 | |
| Proficiency testing and reference materials | 5 | |
| Two- and three-class microbiological sampling plans | 3 | |
| Statistical process control: control charts for variables and attributes | 1 | |
| Raw-material, in-process and finished-product quality control | 1 |
Choose an answer, then open the working. Each explanation says why the right answer is right and why the others are not.
A laboratory plots one fat result per day on an individuals chart: mean 4.0%, mean moving range 0.15%. What are the control limits? (E₂ = 2.66)
B. 3.60% to 4.40%
Limits = X̄ ± 2.66 MR̄ = 4.0 ± 2.66 × 0.15 = 4.0 ± 0.40, i.e. 3.60% to 4.40%. 3.55 to 4.45 uses 3 × MR̄; 3.85 to 4.15 uses ±MR̄; 3.70 to 4.30 uses ±2MR̄.
Further reading: Montgomery, Introduction to Statistical Quality Control; ISO 2859-1
For the same plan (n = 5, c = 2, m = 10², M = 10³ cfu/g), one unit gives 2000 cfu/g and the others are below 100. What is the decision?
B. Reject: one unit exceeds M
Any unit above M makes the lot unacceptable regardless of c. Averaging or retesting to replace a result is not permitted by the plan.
Further reading: ICMSF, Microorganisms in Foods 7; Codex CXG 21-1997
What new requirement did ISO/IEC 17025:2017 add for statements of conformity?
B. A documented decision rule that accounts for uncertainty
When reporting conformity with a specification, the laboratory must define and agree the decision rule, including how measurement uncertainty is taken into account. The other options are not requirements.
Further reading: ISO/IEC 17025:2017; Eurachem guides; ISO 13528
Why is chain-of-custody documentation essential for official food samples?
C. It proves who held the sample and when
Legal samples must be shown to be the sample taken, unaltered; custody records track every transfer. Price, validation and duplicates are separate matters.
Further reading: ISO/IEC 17025:2017; Eurachem guides; ISO 13528
A 100 mL volumetric flask has a stated tolerance of ±0.08 mL with no further information. What is the standard uncertainty from the tolerance?
C. 0.046 mL
With a rectangular distribution, u = a/√3 = 0.08/1.732 = 0.046 mL. 0.080 uses the half-width directly; 0.040 divides by 2; 0.033 divides by √6 (triangular).
Further reading: ISO/IEC 17025:2017; Eurachem guides; ISO 13528
A contaminant has a legal limit of 10 µg/kg. A laboratory reports 11.5 ± 2.0 µg/kg (k = 2). Under the rule "non-compliant if result − U exceeds the limit", what is the decision?
A. Not shown to be non-compliant
11.5 − 2.0 = 9.5, which is not above 10, so non-compliance is not established beyond reasonable doubt under this rule; but neither is compliance proved, since 11.5 + 2.0 exceeds the limit. Retesting until a convenient result appears is not valid.
Further reading: ISO/IEC 17025:2017; Eurachem guides; ISO 13528
What distinguishes a certified reference material from an ordinary reference material?
D. A certificate with value, uncertainty and traceability
A CRM is characterised by a metrologically valid procedure and carries a certificate stating the property value, its uncertainty and traceability. Price and matrix are not defining, and in-house materials are not certified.
Further reading: ISO/IEC 17025:2017; Eurachem guides; ISO 13528
To which standard are producers of reference materials accredited?
A. ISO 17034
ISO 17034 covers reference-material producers; ISO/IEC 17043 covers proficiency-testing providers; ISO/IEC 17020 inspection bodies; ISO 19011 is audit guidance.
Further reading: ISO/IEC 17025:2017; Eurachem guides; ISO 13528
Which Codex guideline recommends accreditation, proficiency testing, validated methods and internal QC for laboratories involved in import and export control of foods?
A. CXG 27-1997
CXG 27-1997 covers assessment of the competence of testing laboratories involved in import and export control. CXG 50-2004 is on sampling, CXG 21-1997 on microbiological criteria and CXC 1-1969 on food hygiene.
Further reading: ISO/IEC 17025:2017; Eurachem guides; ISO 13528
How does accreditation differ from certification?
B. Accreditation: competence; certification: conformity
Accreditation (by a body under ISO/IEC 17011) recognises competence of conformity-assessment bodies such as labs; certification (by a certification body) attests that a system, product or person conforms to requirements. Either may be voluntary or required depending on the context.
Further reading: ISO/IEC 17025:2017; Eurachem guides; ISO 13528